Fluke MET/CAL Procedure
=============================================================================
INSTRUMENT: Beta 230: (1 year) CAL VER /5520
DATE: 02-Sep-97
AUTHOR: Fluke Corporation
REVISION: $Revision: 1.1 $
ADJUSTMENT THRESHOLD: 70%
NUMBER OF TESTS: 41
NUMBER OF LINES: 305
CONFIGURATION: Fluke 5520A
=============================================================================
#
# Compatibility:
# 5500/CAL 5.0 or MET/CAL 5.0 or later.
#
# Subprocedures:
# None
#
# Required Files:
# None
#
# System Specifications:
# TUR calculation is based on specification interval of the accuracy file.
# The default 5520A accuracy file contains 90 day specs.
#
# Fluke makes no warranty, expressed or implied, as to the fitness
# or suitability of this procedure in the customer's application.
#
STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON
1.001 ASK- R N P F V
1.002 ASK+ K X
1.003 HEAD PRELIMINARY INSTRUCTIONS
1.004 DISP Connect UUT to AC power and turn unit ON.
1.004 DISP
1.004 DISP Warm-up time: 30min - REQUIRED
1.004 DISP Temperature: 25 +/-10 degC
1.004 DISP Humidity: 55 +/-10%
1.005 DISP [32] WARNING
1.005 DISP HIGH VOLTAGE is used or exposed during the performance
1.005 DISP of this calibration. DEATH ON CONTACT may result if
1.005 DISP personnel fail to observe safety precautions.
1.006 HEAD {VOLTAGE}
1.007 DISP Connect the 5520A to the UUT as follows:
1.007 DISP [32] 5520A NORMAL HI to UUT + V INPUT
1.007 DISP [32] 5520A NORMAL LO to UUT - V INPUT
1.008 DISP Press button to select input function.
1.008 DISP Press button to select Voltage.
1.009 5520 0.1 0.090000V 0.01% 1e-6U 2W
2.001 5520 1 0.90000V 0.01% 1e-5U 2W
3.001 5520 10 9.0000V 0.01% 1e-4U 2W
4.001 HEAD {mV}
4.002 DISP Press button to select input function.
4.002 DISP Press button to select millivolts.
4.003 5520 100 90.000mV 0.01% 0.001U 2W
5.001 5520 1000 900.00mV 0.01% 0.01U 2W
6.001 5520 10000 9000.0mV 0.01% 0.1U 2W
7.001 HEAD {CURRENT}
7.002 DISP Connect the 5520A to the UUT as follows:
7.002 DISP [32] 5520A AUX HI to UUT + mA INPUT
7.002 DISP [32] 5520A AUX LO to UUT - mA INPUT
7.003 DISP Press button to select input function.
7.003 DISP Press button to select current in mA.
7.004 5520 20 20.000mA 0.015% 0.001U 2W
8.001 5520 50 50.00mA 0.015% 0.01U 2W
9.001 HEAD {RESISTANCE}
9.002 DISP Connect the 5520A to the UUT as follows:
9.002 DISP [32] 5520A AUX HI to UUT 4W HI INPUT
9.002 DISP [32] 5520A AUX LO to UUT 4W LO INPUT
9.002 DISP [32] 5520A NORMAL HI to UUT 2W HI INPUT
9.002 DISP [32] 5520A NORMAL LO to UUT 2W LO INPUT
9.003 DISP Press button to select input function.
9.003 DISP Press button to select Ohms function.
9.003 DISP Select 4W termination operation and press .
9.004 5520 900 400.00Z 0.025/ 0.01U 4W
10.001 5520 900 900.00Z 0.05/ 0.01U 4W
11.001 HEAD {100ohm RTD(385)}
11.002 HEAD 100ohm RTD(385) - {INPUT(read)}
11.003 DISP Press to select an input function.
11.003 DISP Press to select the RTD function.
11.003 DISP Select the 385 type, and press .
11.003 DISP Select degrees Celsius, and press .
11.003 DISP Select 4W termination type, and press .
11.004 5520 800.00degC 0.26U R1 4W
12.001 HEAD {TEMPERATURE (TC)}
12.002 DISP This procedure will test one of the following
12.002 DISP thermocouple types: J,K,T,E,R,S,B, or N.
# C,D, and G type TC not verified 5520A Output N/A).
12.003 MEM2 Which thermocouple type do you wish to test (JKTERSBN)?
12.004 JMPL TYPE_J ZCMPI(MEM2, "J")
12.005 JMPL TYPE_K ZCMPI(MEM2, "K")
12.006 JMPL TYPE_T ZCMPI(MEM2, "T")
12.007 JMPL TYPE_E ZCMPI(MEM2, "E")
12.008 JMPL TYPE_E ZCMPI(MEM2, "R")
12.009 JMPL TYPE_E ZCMPI(MEM2, "S")
12.010 JMPL TYPE_E ZCMPI(MEM2, "B")
12.011 JMPL TYPE_E ZCMPI(MEM2, "N")
12.012 DISP Only the following thermocouples are supported
12.012 DISP by this calibrator: J,K,T,E,R,S,B,N.
12.013 JMP 12.003
12.014 LABEL TYPE_T
12.015 HEAD {T-type THERMOCOUPLE}
12.016 DISP Connect the 5520A TC terminals to the UUT TC INPUT
12.016 DISP using Type T thermocouple wire and connectors.
12.017 DISP Press to select an input function.
12.017 DISP Press to select thermocouple function.
12.017 DISP Select Type T thermocouple, and press .
12.017 DISP Select degrees Celsius, and press .
12.018 HEAD {INPUT(read)}
12.019 M5520 TCREF 21degC
12.020 5520 -250.0degC 2U _T TC 2W
13.001 5520 -200.0degC 1U _T TC 2W
14.001 5520 -100.0degC 0.5U _T TC 2W
15.001 5520 400.0degC 0.3U _T TC 2W
16.001 M5520 *
16.002 HEAD {OUTPUT(source)}
16.003 DISP Change the thermocouple connections on the UUT
16.003 DISP from TC-INPUT to TC-OUTPUT.
16.004 DISP Press